Sodium analyzer CA76NA

CA76NA - Potentiometric sodium analyzer for the monitoring of boiler feedwater, steam, condensate ©Endress+Hauser
Multi-sequencer for up to 6 sample channels ©Endress+Hauser
Flow-through cell with separate pH electrode ©Endress+Hauser
Easy commissioning of the CA76NA sodium analyzer ©Endress+Hauser

Potentiometric system for the monitoring of boiler feedwater, steam, condensate and ion exchangers

Protect your water steam cycles and ion exchangers with the CA76NA sodium analyzer! An increase in sodium levels in high-purity or boiler feedwater indicates the presence of unwanted dissolved impurities. This can cause serious damage and consequently expensive repairs. CA76NA provides highly accurate online monitoring of water quality and enables fast detection of critical events. The sodium analyzer supports your preventive maintenance strategy to keep your power plant running.

  • Benefícios

    • Protection of expensive plant equipments in the power industry: CA76NA sodium analyzer monitors low levels of sodium in water steam cycles detecting carryover and impurities in your power plant.

    • Early indication of condenser leaks via online sodium...

  • Campo de aplicação

    The CA76NA sodium analyzer monitors low levels of sodium in the power industry and utilities:

    • Indication of the purity of water and steam used in boilers and turbines.

    • Provision of early warning for condenser leakages or malfunction of ion exchangers.

    • ...

Características e especificações

Voltar

Details

  • Measuring principle

    Potentiometric / ISE

  • Characteristic

    Analyser for sodium in boiler feedwater, steam, condensate and ion exchangers

  • Size

    850 x 450 x 250 mm
    33.5 x 17.7 x 9.8 inch

  • Design

    Panel stainless steel 1.4301

  • Process temperature

    10 to 40 °C
    (50 to 104 °F)

  • Ambient temperature

    5 to 40 °C
    (41 to 104 °F)

  • Process pressure

    1 to 5 bar (abs)
    14.5 to 72.5 psi

  • Sample flow rate

    10 to 15 l/h
    2.64 to 3.96 gal/h

  • Application

    Online sodium monitoring in boiler feed water, water steam cycles, condensate and ion exchangers

  • Power supply

    100 to 240 V AC, 50/60 Hz

  • Output

    0/4 to 20 mA

  • Measurement range

    0.1 to 9999 µg/l (ppb) Na

  • Consumables

    CA76NA sodium electrode,
    CA76NA pH combination electrode
    CA76NA standard solution 5100 ppb Na

Documentos / Manuais / Software

Acessórios / peças sobressalentes

Nr
Order code
Specification